상세 보기
뇌동맥류 코일 색전술시 발생한 코일의 절단과 치료
초록
목 적 : Following increase of coil embolization, there results were growing better and better in case of intracranial aneurysmal subarachnoid hemorrhage, complications and technical problems were increased, too. So we were reported our unusual experience of coil broken during coil embolization in aneurysmal subarachnoid hemorrhage with literature reviews. 대 상: A 63-aged man was admitted our hospital because of spontaneous subarachnoid hemorrhage due to anterior communicating artery aneurysmal rupture(H-H grade IV and Fisher grade III). During operation, successfully removed intra-aneurysmal coil with piece-meal pattern because of swollen coil surface. The broken site was located just distal to detachemnt zone of coil. 결 과: There were nseveral noted problems, such as device not properly purged of air, coil, coil prolaps into parent vessel, microcatheter tip movement, vessel occlusion, aneurysmal perforation and thromboembolism. But broken of coil was very rare. In reportedcase of SR type of GDC, repeated coil moving or bending was main reason of broken. In our case, the coil broken was developed at first attept to embolization. We were suspected mal-product, intra-microcatheter kinking due to clot and/or microcatheter wedge. 결론:There were highly morbidity and mortality was expected in case of coil broken. We must do careful inspection of microcateter and coil before procedure. In case of coil broken, we recommended cessation of procedure and considered surgical remove, because continuing procedure make high morbidity and mortality.
- 제목
- 뇌동맥류 코일 색전술시 발생한 코일의 절단과 치료
- 제목 (타언어)
- 영문제목
- 저자
- Yoon Seung-Hwan
- 학회명
- 제 45차 대한신경외과학회 추계학술대회