Evaluation of Clinical Risk according to Multi-Leaf Collimator Positioning Error in Spinal Radiosurgery
  • 강동진
  • 오건
  • 신영주
  • 강진규
  • 정재용
  • 외 1명

초록

The purpose of this study is to evaluate the clinical risk of spinal radiosurgery by calculating the dose difference due to dose calculation algorithm and multi-leaf collimator positioning error. The images acquired by the CT simulator were recalculated by correcting the multi-leaf collimator position in the dose verification program created using MATLAB and applying stoichiometric calibration and Monte Carlo algorithm. With multi-leaf collimator positioning error, the clinical target volume (CTV) showed a dose difference of up to 13% in the dose delivered to the 95% volume, while the gross tumor volume (GTV) showed a dose difference of 9%. The average dose delivered to the total volume showed dose variation from –8.9% to 9% and –10.1% to 10.2% for GTV and CTV, respectively. The maximum dose delivered to the total volume of the spinal cord showed a dose difference from –14.2% to 19.6%, and the dose delivered to the 0.35 ㎤ volume showed a dose difference from –15.5% to 19.4%. In future research, automating the linkage between treatment planning systems and dose verification programs would be useful for spinal radiosurgery.

키워드

척추 방사선수술다엽콜리메이터 위치 오차화학 양론적 교정전산모사 알고리즘선량 검증 프로그램Spinal radiosurgeryMulti-leaf collimator positioning errorStoichiometric calibrationMonte-Carlo algorithmDose verification program
제목
척추 방사선수술 시 다엽콜리메이터 위치 오차의 임상적 위험성 평가
제목 (타언어)
Evaluation of Clinical Risk according to Multi-Leaf Collimator Positioning Error in Spinal Radiosurgery
저자
강동진오건신영주강진규정재용이보람
발행일
2023-12
유형
Y
저널명
방사선기술과학
46
6
페이지
527 ~ 533