상세 보기
모니터링 정확도와 운용 강건성을 고려한 개인전투체계용 착용형 생체센서 어레이의 최적 위치 분석
- 하슬기;
- 박상헌;
- 임현철;
- 백승호;
- 김도경;
- ... 윤상희
초록
Monitoring for the physiological state of a solider is essential to the realization of individual combat system. Despite all efforts over the last decades, there is no report to point out the optimal location of the wearable biosensors considering both monitoring accuracy and operational robustness. In response, we quantitatively measure body temperature and heartrate from 34 body parts using 2 kinds of biosensor arrays, each of which consists of a thermocouple(TC) sensor and either a photoplethysmography(PPG) sensor or an electrocardiography(ECG) sensor. The optimal location is determined by scoring each body part in terms of signal intensity, convenience in use, placement durability, and activity impedance. The measurement leads to finding the optimal location of wearable biosensor arrays. Thumb and chest are identified as best body parts for TC/PPG sensors and TC/ECG sensors, respectively. The findings will contribute to the successful development of individual combat system.
키워드
- 제목
- 모니터링 정확도와 운용 강건성을 고려한 개인전투체계용 착용형 생체센서 어레이의 최적 위치 분석
- 제목 (타언어)
- Analysis of the Optimal Location of Wearable Biosensor Arrays for Individual Combat System Considering Both Monitoring
- 저자
- 하슬기; 박상헌; 임현철; 백승호; 김도경; 윤상희
- 발행일
- 2019-04
- 유형
- Y
- 저널명
- 한국군사과학기술학회지
- 권
- 22
- 호
- 2
- 페이지
- 287 ~ 297