Objective: To evaluate the safety and clinical outcomes of the Passeo-18 Lux drug-coated balloon (DCB) in endovascular revascularization procedures under real -world conditions in a Korean population with atherosclerotic disease of the infrainguinal arteries, including below-the-knee (BTK) arteries. Materials and Methods: Eight institutions in the Republic of Korea participated in this prospective, multicenter, single -arm, post-market surveillance study. Two hundred patients with Rutherford class 2-5 peripheral arterial disease and infrainguinal lesions suitable for endovascular treatment were competitively enrolled. Data were collected at baseline, the time of intervention, discharge, and 1-, 6-, 12-, and 24-month follow-up visits. The primary safety endpoint was freedom from major adverse events (MAE) within 6 months (except when limiting the time frame for procedure- or device-related mortality to within 30 days), and the primary effectiveness endpoint was freedom from clinically driven target lesion revascularization (CDTLR) within 12 months after the procedure. Results: A total of 197 patients with 332 target lesions were analyzed. Two-thirds of the patients had diabetes mellitus, and 41.6% had chronic limb-threatening ischemia. The median target lesion length was 100 mm (interquartile range: 56-133 mm). Of the target lesions, 35.2% were occlusions, and 14.8% were located in the BTK arteries. Rate of freedom from MAE was 97.9% at 6 months, and the rate of freedom from CD-TLR was 95.0% and 92.2% at 12 and 24 months, respectively. Subgroup analysis of 43 patients and 49 target lesions involving the BTK arteries showed rate of freedom from MAE of 92.8% at 6 months and rates of freedom from CD-TLR of 88.8% and 84.4% at 12 and 24 months, respectively. Conclusion: The results of the present study, including the BTK subgroup analysis, showed outcomes comparable to those of other DCB studies, confirming the safety and effectiveness of Passeo-18 Lux DCB in the Korean population.

키워드

Peripheral artery diseaseDrug-coated balloonKoreanBelow-the-kneePERIPHERAL ARTERIAL-DISEASEFEMOROPOPLITEAL LESIONSRANDOMIZED-TRIALRISK-FACTORSPACLITAXELANGIOPLASTYRESTENOSISEFFICACYOUTCOMESPREVENTION
제목
Safety and Effectiveness of Passeo-18 Lux Drug-Coated Balloon Catheter in Infrainguinal Endovascular Revascularization in the Korean Population: A Multicenter Post-Market Surveillance Study
저자
Choi, Tae WonWon, Je HwanJae, Hwan JunJeon, Yong SunPark, Sang WooKo, Gi-YoungYim, Nam YeolWon, Jong YunKim, Chang WonKim, Jinoo
DOI
10.3348/kjr.2024.0099
발행일
2024-06
유형
Article
저널명
Korean Journal of Radiology
25
6
페이지
565 ~ 574