Effects of Non-invasive Keratograph Break-Up Time on the Repeatability of Keratometry Measurements

  • Shin, Haeeun
  • Jung, Soyeon
  • Jung, Ji Won
Citations

WEB OF SCIENCE

2
Citations

SCOPUS

2

초록

Purpose: To evaluate the effects of the non-invasive keratograph break-up time on the repeatability of keratometric measure-ments derived using the Keratograph 5M in patients with dry eye syndrome. Methods: Thirty-six patients with dry eye and 30 controls were enrolled. We measured keratometric values twice in all subjects and explored whether the two measurements differed by more than 0.5 diopter (D) (the absolute value). We also evaluated the relationship between the absolute difference and the non-invasive keratograph break-up time (NIKBUT). Results: The intraclass correlation coefficients for astigmatism, flat keratometry (K), steep K, and mean K were all higher than 0.95 in patients, thus lower than in controls. The proportion of eyes with absolute differences over 0.5 D between the two kerato-metric values differed significantly between the dry eye and control groups in terms of the flat K and Kmax values (p = 0.033, 0.037). The average NIKBUT was negatively correlated with the absolute differences in Kmax and astigmatism (p = 0.030, R2 = 0.134 and p = 0.017, R2 = 0.160). Conclusions: In patients with dry eye syndrome, keratometric measurements are reliably repeatable, but slightly less so than in normal controls. The proportions of eyes exhibiting absolute keratometric differences over 0.5 D (two measurements) differed in the dry eye and control groups. The lower the non-invasive keratograph break-up time, the poorer the repeatability of kerato-metric measurements in the dry eye group. Therefore, care is required when obtaining keratometric values for such patients. J Korean Ophthalmol Soc 2023;64(2):99-107

키워드

Dry eye syndromeKeratometryKeratograph 5MNoninvasive keratograph break-up timeINTRAOCULAR-LENS POWERFILM
제목
Effects of Non-invasive Keratograph Break-Up Time on the Repeatability of Keratometry Measurements
저자
Shin, HaeeunJung, SoyeonJung, Ji Won
DOI
10.3341/jkos.2023.64.2.99
발행일
2023-02
유형
Article
저널명
대한안과학회지
64
2
페이지
99 ~ 107