Trajectory of bolt and length of plate in femoral neck system determine the stability of femur neck fracture and risk of subsequent subtrochanteric fracture : a finite element analysis

  • Jung, Chang-Ho
  • Cha, Yonghan
  • Chung, Jun Young
  • Park, Chan Ho
  • Kim, Tae Young
  • ... Yoo, Jun-Il
  • 외 2명
Citations

WEB OF SCIENCE

8
Citations

SCOPUS

11

초록

BackgroundThis study aimed to analyze the differences in the stability of fractures, stress distribution around the distal-most screw according to the length of the plate and the trajectory of the bolt in Pauwels type III femoral neck fracture using the femoral neck system (FNS).MethodsFinite element models of Pauwels type III femoral neck fractures were established with surgical variations in the trajectory of the bolt (central, inferior, valgus, and varus) and length of the lateral plate (1- and 2-hole plate). The models were subsequently subjected to normal walking and stair-climbing loads.ResultsThe screw-holding cortical bone in subtrochanter in the model with a 2-hole plate and the bolt in the inferior trajectory and the models with 1-hole or 2-hole plate and the bolt in valgus trajectory had shown greater maximum principal strain than the models with central or varus trajectories. The gap and sliding distance on the fracture surface were larger with inferior or varus trajectories of the bolt and smaller with the valgus trajectory of the bolt under both loads, compared to those of the central trajectory.ConclusionFor the fixation of Pauwels type III femoral neck fracture, the trajectory of the FNS bolt and the length of the plate affect the mechanical stability of the fracture and the strain of cortical bone around the distal-most screw. The surgical target should stay on the central trajectory of the bolt and the 2-hole plate's mechanical benefits did not exceed the risk.

키워드

Bolt trajectoryFemoral neck fracturesFixationPauwels type IIISUITABLE FNS SPECIFICATIONDYNAMIC HIP SCREWBIOMECHANICAL EVALUATIONINTRACAPSULAR FRACTURESPAUWELS CLASSIFICATIONINTERNAL-FIXATIONYOUNG-PATIENTSMANAGEMENTCHOOSEJOINT
제목
Trajectory of bolt and length of plate in femoral neck system determine the stability of femur neck fracture and risk of subsequent subtrochanteric fracture : a finite element analysis
저자
Jung, Chang-HoCha, YonghanChung, Jun YoungPark, Chan HoKim, Tae YoungYoo, Jun-IlKim, Jung-TaekJeon, Yongho
DOI
10.1186/s12891-023-06579-4
발행일
2023-06-07
유형
Article
저널명
BMC Musculoskeletal Disorders
24
1